久久伊人五月丁香狠狠色,人妻少妇精品视频一区二区三区,AV无码AV高潮AV喷吹免费,男JI大巴进入女人的视频小说

EDS (Energy Dispersive Spectrometer)

SEM/EDS- Scanning Electron Microscopy/ Energy Dispersive Spectroscopy

Scanning electron microscopy (SEM) is a method for high resolution surface imaging. The SEM uses electron for imaging. The advantages of SEM over light microscopy include greater magnification and much greater depth of field. Different elements and surface topography emit different amounts of electrons; due to this the contrast in the electron micrograph (picture) is representative of surface topography and composition.

Energy dispersive spectroscopy (EDS) measures the number of x-rays produced by a solid sample when irradiated by electrons versus the energy of these x-rays. The EDS technique identifies and quantifies the element constituents of the sample.

Technical Capabilities

 

Signal Detected: SEI/ BSE/ Characteristic x-rays

Elements Detected: B-U

Detection Limits: 0.1 – 1 at%

Depth Resolution: 0.5 – 3 μm

 

Applications for SEM/EDS Analysis

 

• Study fracture surfaces -both metal and polymer

• Study microstructure of a sample

• Obtain elemental information on surface

• Identify foreign materials (particles/ contaminants/ corrosions)

• Determine the thickness & composition of coatings

 

Relevant Industries for Auger Analysis

  • Aerospace
  • Automotive
  • Biomedical/biotechnology
  • Compound Semiconductor
  • Data Storage
  • Defense
  • Displays
  • Electronics
  • Industrial Products
  • Lighting
  • Pharmaceutical 
  • Photonics
  • Polymer
  • Semiconductor
  • Solar Photovoltaic
  • Telecommunications

Strengths

  • Quick, “first look” analysis
  • Versatile, inexpensive, and widely available
  • Quantitative for some samples (flat, polished, homogeneous)
  • Small area analysis (as small as 30 nanometers)
  • Semi-quantitatively analysis
  • Excellent surface sensitivity
  • Good depth resolution

Limitations

  • Semi-quantification for samples that are not flat, polished, and homogeneous
  • Size restrictions on samples
  • Samples must be vacuum compatible (not ideal for organic material)
  • Analysis (or coating) may influence subsequent surface analysis
  • Limited sensitivity for low-Z elements
Copyright © 2010 cyu-china.com.cn All Rights Reserved .  
国产肉体xxxx裸体137大胆| 亚洲高清专区日韩精品| 人妻夜夜爽天天爽一区| 亚洲精品无码午夜福利中文字幕| 国产人妻精品一区二区三区不卡| 日日躁夜夜躁狠狠躁超碰97|